Clarity Octopus EMG/NCS/EP System | 2 Channel | 16-Bit A/D Converter
The Clarity Octopus 2-Channel is a fully computerized, PC-based electromyography (EMG), nerve conduction study (NCS), and evoked potential (EP) system configured on a dedicated mobile workstation cart.
Product description
Key Features:
Dual-channel miniature module with extremely low noise amplifiers
Powerful integrated electrical stimulator with both current and voltage modes
Auditory, visual, and photic stimulators for evoked potential testing
PC-based software stores data and prints reports from any Windows-compatible printer
Sturdy mobile cart with arm for easy positioning
Technical Specifications:
Channels: 2 (expandable platform, up to 8)
Sensitivity: 0.1 µV/div to 20 mV/div
Sweep speed: 1 ms/div to 1000 ms/div
Gain: 50 to 20,000, with fine control
Noise level: <0.4 µV RMS (100 Hz–10 kHz)
Input impedance: >10 MΩ; CMRR: >100 dB
Filters: high 100 Hz–10 kHz, low DC–500 Hz
A/D conversion: 16-bit SAR, sampling up to 80 kHz
Electrical stimulator: pulse width 20–1000 µs, rate 0.1–30 Hz; current mode 0–100 mA, voltage mode 50–400 V
Auditory stimulator: click/tone-pip/tone-burst, 0–100 dB nHL, 250 Hz–7.5 kHz
Tests supported: nerve conduction (MNC, SNC, F-wave, H-reflex, blink reflex, RNS), EMG (spontaneous, quantitative, interference pattern, single fibre), evoked potentials (BAEP, VEP, SEP, MLR, LLR, P300), HRV, SSR